PMX3300 Series Continuous-Intermittent 3-Mode

$0.00

SKU: PMX3300 Series Continuous-Intermittent 3 Mode Categories: ,

Product Description

The 3-Mode Continuous & Intermittent model series provides reliable and accurate cycling of the regulator On and Off during the intermittent suctioning. The 3-Mode Continuous & Intermittent series has a peak vacuum limit feature to prevent adjusting the vacuum too high. This series is ideal for sensitive applications where limiting Max vacuum is critically important. The intermittent timing cycle is factory set at approximately 16 seconds on and 8 seconds off.

Features:

  • Factory preset timing cycles set at approximately 16 seconds On, 8 seconds Off
  • Fully adjustable, 2-to-1 ratio timing cycles
  • Modular components are attached to a cast metal back plate, simplifying service
  • Available in digital and analog gauges
  • Low to high vacuum rates
  • Large, easy-to-grip knobs

Additional Information

Gauge Type

Analog, Digital

Regulator Mode

Continuous-Intermittent 3 Mode (0-200 mmHg), Continuous-Intermittent 3 Mode (0-300 mmHg)

Vacuum Source Connection

1/8" NPT Female, DISS Female Hex Nut, DISS Female Hand Tight, Ohmeda Quick Connect, Chemetron Quick Connect, Oxequip Quick Connect, Medstar Quick Connect, Puritan-Bennett Quick Connect, Schrader Quick Connect, Tubing Nipple, 90⁰ Tubing Nipple, 90° Tubing Nipple

Patient Connection

1/8" NPT Female, Tubing Nipple, DISS Male, Ohmeda Style Trap Bottle Connection, Filtered Vac Trap

Vac Trap

No, Yes – Reusable Vac Trap, Yes – Filtered Vac Trap

Dimensions
Depth: 4.5 in (11.4 cm)
Width: 3.8 in (9.5 cm)
Height: 5.7 in (14.4 cm)
Weight: 1.1 lbs (.50 kg)
Gauge Range
PMX3300: 0 – 200 mmHg – Full Vacuum*PMX3300E: 0 – 200 mmHg (0 – 26 kPa)
*Counterclockwise direction

Gauge Accuracy
Analog: ± 5% of MAXDigital Gauge: ± 1% of Full Scale
Analog Gauge: ± 5% of MAX within ref. Indicator

Vacuum Ports
1/8″ NPT Female
Modes
REG. (Regulated): provides an adjustable, continuous vacuum levelOFF: No Vacuum

INT. (Intermittent): provides an adjustable vacuum level that cycles between ON and OFF

Flow Rates
REG: 50 l/min
INT: 10 l/min
*Maximum flow is obtained with a vacuum source of 21″ Hg (71.1 kPa)
Maximum Vacuum
REG. Mode @ Max – min. 100 mmHg above calibrated range
Intermittent Cycle Time
Factory set at sixteen (16) seconds ON and eight (8) seconds OFF (Reference only)
Operating Environmental Limits
0°F to 122°F (-18°C to 50°C)
Recommended Environmental Operating Limits
55°F to 85°F (13°C to 29°C)
Storage Environmental Limits
Temperature Range:
-4°F to 140°F (-20°C to 60°C)Humidity:
Max 95% Non-condensing

Battery
Digital Vacuum Gauge Models ONLY
3 Volt Lithium, ½ AA

Manuals

PMX Large Gauge Vacuum Regulator User Manual – English

Brochures

PMX Series Overview
Product Catalog – Vacuum Regulators

LIMITED WARRANTY AND LIMITATION OF LIABILITY

Precision Medical, Inc. warrants that the Large Gauge Medical Vacuum Regulator (the Product) will be free of defects in workmanship and/or material for the following period:

Ten (10) years from date of shipment.

Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.

ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.

The representative of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.

THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.

Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of nonconformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.

Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.